Development and Validation of a Continuous Real-Time Optical Sensor for Indocyanine Green Clearance Measurement During Ex-Vivo Perfusion of Human Livers
Edmund N.J. Derwent, Charles W.G. RisbeyiD, Anita NiuiD, Paul YousifiD, Nidula Fonseka, Spencer Curry, Chloe Kwok-Long SeowiD, Isaac Ng, Geoffrey William McCaughaniD, Michael Crawford, Carlo PulitanòiD, Daniel BabekuhliD
Abstract
Liver transplantation remains the only curative treatment for end-stage liver failure, yet its impact is constrained by organ shortages and graft non-utilisation. Machine perfusion (MP) enables ex-vivo assessment of donated livers; however, existing viability criteria rely on intermittent sampling, reducing temporal resolution and accuracy. Indocyanine green (ICG), a clinically validated dye cleared exclusively by hepatocytes, provides a continuous index of hepatic function beyond initial injury. Accordingly, we present a non-invasive, clamp-on optical sensor that enables continuous, real-time quantification of ICG clearance during MP. The sensor consists of a clamp-on module with an 808nm laser and phototransistor connected to a microcontroller-based unit and computer for real-time plotting. The raw phototransistor signal was linearised to a unitless absorbance signal proportional to perfusate ICG; bi-exponential fitting yielded plasma disappearance rate (PDRbi, %/min) and the 15-minute residual fraction (R15). Across 10 whole and 3 split human livers (45 boluses; 13 paired with spectrophotometry), the sensor closely matched spectrophotometric measurements (mean R2 = 0.994; range 0.983-0.999). The sensor resolved expected physiological trends: ICG clearance increased with temperature (PDRbi: 8.2%/min (subnormothermic MP) to 22.6%/min (normothermic MP) (n=4); 9.3%/min (32{degrees}C) to 11.9%/min (36{degrees}C) (n=1)). The sensors continuous signal traces also revealed early mixing dynamics and medication-related effects that are missed by intermittent sampling. This optical sensor enables accurate, real-time monitoring of ICG clearance during ex-vivo perfusion. The ex-vivo setting is uniquely positioned to validate ICG clearance models and enhance clinical interpretation.
- Citations
- 0
- FWCI
- 0.00
- Field percentile
- 15th percentile
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Real time evaluation of the liver microcirculation by whole organ machine perfusion within an MRI system
Zainab L. Rai, Natalie Holroyd, … Brian Ritchie Davidson·NMR in BiomedicinePeer-reviewed·Feb 6, 2026
- Prediction of survival after partial hepatectomy using a physiologically based pharmacokinetic model of indocyanine green liver function tests
A. Köller, J. Grzegorzewski, … M. König·bioRxiv·Jun 15, 2021
- Physiologically based modeling of the effect of physiological and anthropometric variability on indocyanine green based liver function tests
Adrian Köller, Jan Grzegorzewski, Matthias König·Frontiers in PhysiologyPeer-reviewed·Nov 22, 2021
- Improving retention time of Indocyanine Green for in vivo two-photon microscopy using liposomal encapsulation
Alankrit Tomar, Noah Stern, … Andrew K. Dunn·bioRxiv·Mar 20, 2025
- Monitoring deep-tissue oxygenation with a millimeter-scale ultrasonic implant
S. Sonmezoglu, J. R. Fineman, … M. M. Maharbiz·bioRxiv·Mar 5, 2021
- A MICROFLUIDIC DEVICE FOR LONG-TERM MAINTENANCE OF ORGANOTYPIC LIVER CULTURES
J. M. de Hoyos-Vega, H. Hong, … A. Revzin·bioRxiv·Jun 18, 2022