Impact of tissue staining and scanner variation on the performance of pathology foundation models: a study of sarcomas and their mimics
Binghao ChaiiD, Jianan CheniD, Paul CooliD, Fatine OumliliD, Anna Tollit, David F. SteineriD, Tapabrata ChakrabortiiD, Adrienne Margaret FlanaganiD
Abstract
Histopathological analysis is considered the gold standard for the diagnosis and prognostication of cancer. Recent advances in AI, driven by large-scale digitisation and pan-cancer foundation models, are opening new opportunities for clinical integration. However, it remains unclear how robust these foundation models are to real-world sources of variability, particularly in H&E staining and scanning protocols. In this study, we use soft tissue tumours, a rare and morphologically diverse tumour type, as a challenging test case to systematically investigate the colour-related robustness and generalisability of seven AI models. Controlled staining and scanning experiments were utilised to assess model performance across diverse real-world data sources. Foundation models, particularly UNI-v2, Virchow and TITAN, demonstrated encouraging robustness to staining and scanning variation, particularly when a small number of stain-varied slides were included in the training loop, highlighting their potential as adaptable and data-efficient tools for real-world digital pathology workflows.
- organism: human
- assay: imaging
- research focus: benchmarking
- research focus: database / resource
- research focus: cell type annotation
- research focus: clustering
- approach: transformer / foundation model
- approach: deep learning
- approach: benchmark study
- approach: experimental assay
In-depth summary
Claude Opus 5.5 · xhigh effort · 8 Nov 2026 · from the full text of version 2 (28 Aug 2025) · CC BY 4.0
Across 14 sarcoma and mimic subtypes, six pathology foundation models beat ResNet50 but lost accuracy under multi-institutional staining; adding 121 externally stained slides narrowed the gap.
Clinical pathology AI must perform consistently across H&E staining protocols and slide scanners, yet stain normalisation can distort morphology, GAN-based correction needs large data and transfers poorly, and foundation-model robustness to these factors is underexplored. Using rare, morphologically diverse soft tissue tumours, the study quantifies that robustness via controlled stain and scanner experiments and releases a multi-lab, multi-scanner slide collection.
Key results
ShownAll six foundation models outperformed a supervised ResNet50 on slides from one lab and one scanner. Every model, however, lost significant accuracy on a real-world test set stained in many hospitals, and ResNet50 degraded most.
Class-wise accuracy, Cohort 1-Test vs Cohort 4-Test: TITAN 0.932 vs 0.869, UNI-v2 0.939 vs 0.842, Virchow 0.921 vs 0.859, CONCH-v1.5 0.911 vs 0.737, PRISM 0.882 vs 0.811, ResNet50 0.821 vs 0.466; chi-squared decline significant for all seven models (TITAN p = 0.004, Virchow p = 0.006, others p < 0.001); best models dropped 6–10% (Discussion). · 2.1 Evaluation of foundation models for…; 3 Discussion; Figure legends
ShownAdding a small set of externally stained slides to the single-lab training data significantly improved real-world accuracy for several patch-level models. It also lifted the weakest classes (nodular fasciitis, intramuscular myxoma), with a minor decline on the in-house test set for some models.
Adding 121 Cohort 4-Train WSIs to Cohort 1-Train (Combined cohort) significantly raised Cohort 4-Test accuracy for Path Foundation (p < 0.001), UNI-v2 (p < 0.002), CONCH-v1.5 (p < 0.001) and ResNet50 (p < 0.001); not significant for Virchow (p = 0.478), TITAN (p = 0.238) or PRISM (p = 0.329); chi-squared tests. · 2.1 Evaluation of foundation models for…; 4.5 Statistical analysis; Figure legends
SuggestedWith staining held constant, CONCH-v1.5, Virchow, PRISM and ResNet50 lost accuracy when trained on Hamamatsu or 3D-Histech scans and tested on Aperio scans. TITAN and UNI-v2 stayed consistently high across all three scanners.
Cohort 3 (335 WSIs; eight cases per diagnosis stained at Lab A, each scanned on Aperio GT 450, NanoZoomer S360 and Pannoramic 1000); one classifier per scanner per model, all tested on Aperio-scanned Cohort 1-Test. Accuracies are shown only graphically; no values or tests are given in the text. · 2.3 How robust are foundation models to scanning…; 4.1 WSI datasets; Figure legends
SuggestedWith very few training slides (1 to 8 per diagnosis), the slide encoder TITAN already performed well from a single slide. Patch encoders such as UNI-v2 started lower but improved quickly as slides were added, and all foundation models beat ResNet50.
Few-shot classifiers trained on 1, 2, 4, 6 and 8 Cohort 3 slides per diagnosis from each scanner, tested on Cohort 1-Test: TITAN exceeded 0.75 accuracy with one case per class; most foundation models exceeded 0.8 accuracy with 8 cases per subtype; best results when the training scanner matched the Aperio test scanner. · 2.4 Do foundation models enable few-shot…; 3 Discussion; Figure legends
SuggestedColour variation followed the scanner more than the staining lab. Stain-vector PCA grouped by scanner, and UNI-v2 tile embeddings formed scanner sub-clusters within each diagnosis. TITAN slide embeddings showed little scanner separation, and neither model grouped by staining lab.
Vahadane stain vectors from randomly sampled 3000 × 3000 pixel tiles of Cohort 2 slides (stained by 30 labs; Fig. 4 illustrates a synovial sarcoma case): Aperio formed a tight cluster, 3D-Histech was widely dispersed and Hamamatsu was intermediate. t-SNE used three random tumour tiles per slide for UNI-v2 and one embedding per slide for TITAN. Both analyses are qualitative only. · 2.2 Impact of staining and scanning variation; Appendix A Extended Data: Embedding space…; Figure legends
Methods, data and reproducibility
WSIs were tiled with TRIDENT at 20x. Patch embeddings (Path Foundation, UNI-v2, CONCH-v1.5, Virchow, ResNet50) fed an attention-based multi-instance learning classifier, and TITAN/PRISM slide embeddings fed logistic regression. Models diagnosed 14 classes under controlled stain, scanner, mixed-training and few-shot experiments.
- Cohort 1: H&E WSIs stained at Lab A and scanned on Leica Aperio GT 450 at 40× equivalent; split 60%/20%/20% for train/validation/test · human (patient tumour cases) · n = 1,394 WSIs (approximately 100 cases per diagnosis)
- Cohort 2: serial 4 µm sections of one representative case per diagnosis, H&E-stained by 30 NHS labs, each slide scanned on three scanners · human (patient tumour cases) · n = 1,243 WSIs
- Cohort 3: eight cases per diagnosis stained at Lab A, each scanned on Aperio GT 450, Hamamatsu NanoZoomer S360 and 3D-Histech Pannoramic 1000 · human (patient tumour cases) · n = 335 WSIs
- Cohort 4: real-world archive slides stained at Lab A or 73 other hospitals and scanned on one platform; split into a Combined-cohort training subset and a test subset · human (patient tumour cases) · n = 934 WSIs (121 for Combined training, 813 remaining)
- Methods & tools
- TRIDENT (tissue detection and tiling), Google Path Foundation, UNI-v2, CONCH-v1.5, TITAN (slide encoder), Virchow, PRISM (slide encoder), ResNet50 CNN (supervised baseline), Attention-based multi-instance learning classifier, scikit-learn logistic regression, Vahadane stain separation with PCA, t-SNE (scikit-learn, Plotly)
- Data & accessions
- https://cbhindex.github.io/visualise_scan_stain_efffects
- Reproducibility gaps
- The Data availability section gives a placeholder ('LINK') instead of a URL for the 1,243-WSI multi-institutional release. Sharing of Cohorts 1, 3 and 4 is not described, and no software versions or hardware are reported.
Limitations (6)
- Only three scanner models (Aperio, Hamamatsu, 3D-Histech) were tested. Devices from other vendors, such as Philips and Roche, were not included, so generalisation to other scanners is not established. · stated by the authors
- Evaluation covered a single task, soft-tissue tumour subtyping. Other tumour types and tasks, such as mitosis detection or genomic marker prediction, were not tested. · stated by the authors
- The stain-vector PCA was too widely spread to quantify how much staining and scanning each contribute to colour variation. · stated by the authors
- Non-Aperio scanned slides were in limited supply, so the scanner experiment trained one classifier per scanner on eight Lab A-stained cases per diagnosis but tested only on Aperio-scanned slides. · stated by the authors
- Patch encoders used an attention-based MIL head while TITAN and PRISM used unregularised logistic regression. Comparisons between slide and patch encoders, including TITAN's few-shot advantage, therefore mix encoder quality with classifier design, and no repeated few-shot draws are reported. · noted in this summary
- Cohort 4 counts are internally inconsistent (934 WSIs vs 105 Lab A-stained plus 838 external), and the 813-slide subset is described as training data although the design implies it is Cohort 4-Test. Scanner and few-shot results lack per-model accuracy values or statistical tests in the text. · noted in this summary
Read this if you deploy or fine-tune pathology foundation models across hospitals and need evidence on how H&E stain variation, scanner variation and few-shot training affect slide-level tumour subtyping.
- Citations
- 6
- FWCI
- 20.37
- Field percentile
- top 1%
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Stain Normalisation Does Not Consistently Improve Matched-Location Retrieval with Pathology Foundation Models
Yang Meng, Binghao Chai, Tapabrata Chakraborti·bioRxiv·Oct 19, 2026
- Understanding Foundation Models in Digital Pathology: Performance, Trade-offs, and Model-Selection Recommendations
Danial Maleki, Nazim Shaikh, … Uday Kurkure·bioRxiv·Sep 12, 2025
- Machine-learning convergent melanocytic morphology despite noisy archival slides
Mikio Tada, Garrett Gaskins, … Elizabeth Keiser·Cell Reports MethodsPeer-reviewed·Oct 1, 2025
- stainlib: a python library for augmentation and normalization of histopathology H&E images
Sebastian Otálora, Niccolò Marini, … Manfredo Atzori·bioRxiv·May 17, 2022
- Examining Batch Effect in Histopathology as a Distributionally Robust Optimization Problem
S. N. Hari, J. Nyman, … E. M. Van Allen·bioRxiv·Sep 14, 2021
- Benchmarking artificial intelligence methods for end-to-end computational pathology
Narmin Ghaffari Laleh, Hannah Sophie Muti, … Jakob Nikolas Kather·bioRxiv·Aug 9, 2021