Performances of NeuMoDx, a random-access system for HBV-DNA and HCV-RNA quantification
J. Besombes, C. Pronier, C. Lefevre, G. Lagathu, A. Maillard, C. Grolhier, V. Thibault
Abstract
Viral loads (VL) monitoring for hepatitis B and C is essential to evaluate disease progression and treatment response. Automated, random-access rapid systems are becoming standard to provide reliable VL to clinicians. The aim of this study was to evaluate the analytical performances of the recently launched NeuMoDx for HBV-DNA and HCV-RNA quantification. Clinical samples routinely quantified on the Beckman-Veris system were either retrospectively (frozen samples; HBV n=178, HCV n=249), or in parallel (fresh primary tubes; HBV n=103, HCV n=124) tested using NeuMoDx. Linearity range was assessed on serial dilutions of high tittered plasmas containing different genotypes for HBV (A-E, n=10) and HCV (1a-b, 2-5, n=12). Overall test failure, mostly internal control amplification failure, was 2.3% and was not influenced by matrix types. For HBV-VL, Kappa agreement was 74%, with 27 (12.6%) discrepancies. Correlation between HBV assays on 72 quantified samples by both methods was excellent (r=0.963) with a mean bias (NeuMoDx-Veris) of 0.21 log IU/mL. For HCV-VL, Kappa agreement reached 94%, with 9 (2.8%) discrepancies. The r-correlation factor between assays on 104 samples was 0.960 with a mean bias of -0.14 log IU/mL (NeuMoDx-Veris). Serial dilutions confirmed the claimed linear ranges for all HBV and HCV genotypes. The mean turnaround time was 72 [55-101] for HBV and 96 [78-133] for HCV. These results obtained on the NeuMoDx confirmed the overall good functionality of the system with a short turn-around-time, full traceability and easy handling. These results on HBV- and HCV-VL look promising and should be challenged with further comparisons.
02 /Specification & Classification
Click any tag to browse matching papers- Research Focus
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Performance of the cobas(R) HBV RNA Automated Investigational Assay for the Detection and Quantification of Circulating HBV RNA in Chronic HBV Patients
C. SCHOLTES, A. T. Hamilton, … F. Zoulim·bioRxiv·Mar 9, 2022
- Multi-site Performance Evaluation of the cobas 5800 System and Comparison to the cobas 6800/8800 Systems for Quantitative Measurement of HBV, HCV, and HIV-1 Viral Load.
J. Tschape, A. Cobernuss-Rahn, … P. Gohl·bioRxiv·Aug 9, 2022
- Systematic review of the validity of the hepatitis B virus detection tests in blood banks from 2000 to 2018.
C. P. Orrego-Marin, J. A. Cardona-Arias, … J. C. Zapata·bioRxiv·Apr 17, 2020
- CCQM-P199: Interlaboratory comparability study of HIV-1 RNA copy number quantification
Alison S. Devonshire, Eloise J. Busby, … Jim F. Huggett·bioRxiv·Apr 12, 2024
- A multiscale model of the action of a capsid assembly modulator for the treatment of chronic hepatitis B
Sarafa A. Iyaniwura, Tyler Cassidy, … Alan S. Perelson·PLoS Computational BiologyPeer-reviewed·May 6, 2025
- Comparison of simple RNA extraction methods for molecular diagnosis of hepatitis C virus in plasma
Sayamon Hongjaisee, Yosita Jabjainai, … Woottichai Khamduang·DiagnosticsPeer-reviewed·Jun 30, 2022