Second-generation dual-channel visible light optical coherence tomography enables wide-field, full-range, and shot-noise limited retinal imaging

J. Wang, S. Nolen, W. Song, W. Shao, W. Yi, J. Yi

Abstract

Visible light optical coherence tomography (VIS-OCT) is an emerging ophthalmic imaging method uniquely featured by ultrahigh depth resolution, retinal microvascular oximetry, and distinct scattering contrast in the visible spectral range. However, the clinical utility of VIS-OCT is impeded by the fundamental trade-off between the imaging depth range and axial resolution, determined by the spectral resolution and bandwidth respectively. While the full potential of VIS-OCT is leveraged by a broad bandwidth, the imaging depth is inversely sacrificed. The effective depth range is further limited by the wavelength-dependent roll-off that the signal-to-noise ratio (SNR) reduces in the deeper imaging range, more so in shorter wavelength. To address this trade-off, we developed a second-generation dual-channel VIS-OCT system including the first linear-in-k VIS-OCT spectrometer, reference pathlength modulation, and per A-line noise cancellation. All combined, we have achieved 7.2dB roll-off over the full 1.74 mm depth range (water) with shot-noise limited performance. The system uniquely enables >60{degrees} wide-field imaging over large retinal curvature at peripheral retina and optic nerve head, as well as high-definition imaging at ultrahigh 1.3 um depth resolution (water). The dual-channel design includes a conventional near infrared (NIR) channel, compatible with Doppler OCT and OCT angiography (OCTA). The comprehensive structure-function measurement by 2nd-Gen VIS-OCT system is a significant advance towards broader adaptation of VIS-OCT in clinical applications.

PDFDOI

02 /Specification & Classification

Click any tag to browse matching papers

Reproducibility ledger

Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.

Sign in to file a reproducibility report.